CVE-2025-27714 is an arbitrary file upload vulnerability [CWE-434] disclosed through CISA ICS Medical Advisory ICSMA-25-100-01. An authenticated attacker can upload arbitrary files through a specific endpoint. Successful exploitation leads to unauthorized remote code execution or system compromise.
The issue affects a medical device product referenced in the CISA advisory. The vulnerability requires network access and low-level privileges, with no user interaction. Because the product operates in clinical environments, exploitation carries operational and patient-safety implications beyond typical IT impact.
Critical Impact
An authenticated attacker on the network can upload malicious files to a vulnerable endpoint and achieve remote code execution on the affected medical device.
Affected Products
- Medical device referenced in CISA ICS Medical Advisory ICSMA-25-100-01
- Specific product and version details are enumerated in the CISA advisory
- No CPE identifiers are published in the NVD record
Discovery Timeline
- 2025-08-21 - CVE-2025-27714 published to the National Vulnerability Database
- 2026-06-17 - Last updated in NVD database
Technical Details for CVE-2025-27714
Vulnerability Analysis
The vulnerability is classified as Unrestricted Upload of File with Dangerous Type [CWE-434]. The affected endpoint accepts file uploads without adequately validating file type, extension, content, or destination path. An attacker authenticated to the device can submit crafted files that the underlying platform later executes or interprets.
Once a malicious payload lands in a web-accessible or interpreter-reachable location, the attacker can trigger it to run arbitrary code in the process context of the receiving service. This yields interactive control of the medical device, exposure of stored data, and a foothold for lateral movement across the clinical network.
The published EPSS probability is low, indicating limited observed exploitation activity to date. However, medical devices typically operate on long patch cycles, so exposure windows can remain open for extended periods.
Root Cause
The root cause is missing or insufficient validation on an upload endpoint. Contributing factors typically associated with [CWE-434] include:
- Failure to enforce an allowlist of file extensions and MIME types
- No server-side content inspection of the uploaded bytes
- Storage of uploaded files inside a directory served or executed by the application
- Predictable or attacker-controlled destination filenames
Attack Vector
The attack vector is network-based and requires authenticated access to the vulnerable endpoint. The attacker submits an HTTP request containing a file payload of a dangerous type, such as a server-side script or executable. The device stores the file and the attacker requests it, causing the runtime to interpret the payload.
No verified proof-of-concept code is available in public sources. Consult the CISA ICS Medical Advisory ICSMA-25-100-01 for vendor-specific technical detail.
Stop Remote Code Execution Before It Runs
Singularity™ Endpoint detects and blocks exploit attempts at machine speed, including zero-days and novel attack techniques.
Stop Exploit AttemptsDetection Methods for CVE-2025-27714
Indicators of Compromise
- Unexpected files with executable or script extensions (.php, .jsp, .aspx, .sh, .py) written to upload directories on the affected device
- HTTP POST requests to file-upload endpoints originating from unusual internal or external source addresses
- New child processes spawned by the web or application service account shortly after an upload request
- Outbound network connections from the device to attacker-controlled infrastructure following an upload event
Detection Strategies
- Inspect web server and application logs for POST requests to upload endpoints with abnormal Content-Type values or oversized payloads
- Monitor filesystem changes in web-accessible directories using file integrity monitoring
- Correlate upload events with subsequent process creation on the device host
- Apply network intrusion detection signatures for file-upload payloads targeting the vulnerable endpoint
Monitoring Recommendations
- Forward device access logs, application logs, and process telemetry to a centralized SIEM for correlation
- Baseline normal upload behavior for the device and alert on deviations in volume, source, or file type
- Segment medical devices from general user networks and monitor east-west traffic reaching upload endpoints
How to Mitigate CVE-2025-27714
Immediate Actions Required
- Review CISA ICS Medical Advisory ICSMA-25-100-01 to identify affected models and firmware versions in your environment
- Restrict network access to the affected devices so only authorized clinical systems can reach management and upload endpoints
- Rotate credentials used to authenticate to the vulnerable device and enforce strong authentication where supported
- Audit upload directories for unexpected files and remove any that cannot be attributed to legitimate operational use
Patch Information
Refer to the vendor mitigation guidance published in CISA ICS Medical Advisory ICSMA-25-100-01. Apply vendor-supplied firmware or software updates according to the vendor's clinical validation and change-management procedures. No independent vendor advisory URL is published in the NVD record beyond the CISA reference.
Workarounds
- Place the affected devices behind a network firewall that blocks untrusted access to the upload endpoint
- Disable or restrict the vulnerable upload function through vendor-supported configuration where feasible
- Require VPN or jump-host access for any administrative interaction with the device
- Enforce least privilege on service accounts and application users that can authenticate to the device